SECTION 5 FEATURES AND CAPABILITIES
______________________________________________________________________
5-6
Sensor information is also stored and can be viewed in the Sensor Data Display screen.
When the memory is full, the oldest data are automatically erased as new data are
stored. Memory will be retained even if the battery has been discharged and remains
when the monitor is in the power off condition.
5.2.22 Saved Settings
After power up, your instrument will revert to user settings that are set up, then saved in
the Advanced Setup Menu.
5.2.23 Battery Operation
In the event of a power failure or interruption of power during a procedure, the monitor
automatically switches to back-up battery operation. During battery operation, a
BATTERY symbol appears in the BIS numeric region of the display. A fully charged
battery will provide a minimum 20 minutes of operation. The thermal printer is disabled
during battery operation. The battery symbol blinks to alert the user when only a few
minutes of back-up power remain. The Save Settings feature is disabled when the
battery symbol is blinking. Battery recharge time is approximately 4 hours with the
monitor ON.
Note:
The monitor must be powered ON to recharge the battery.
5.2.24 Data Transfer and Software Updates
Data may be transferred to and from the A-2000 monitor using the Serial port on the
back panel. Request the Aspect A-2000 Serial Port Specification (P/N 070-0017) for a
complete description of the protocols. This port is also used to easily upgrade system
software.
WARNINGS!
THE USE OF ACCESSORY EQUIPMENT NOT COMPLYING WITH THE
EQUIVALENT SAFETY REQUIREMENTS OF THIS EQUIPMENT MAY
LEAD TO A REDUCED LEVEL OF SAFETY OF THE RESULTING
SYSTEM. CONSIDERATION RELATING TO THE CHOICE SHALL
INCLUDE:
USE OF THE ACCESSORY IN THE PATIENT VICINITY
EVIDENCE THAT THE SAFETY CERTIFICATION OF THE
ACCESSORY HAS BEEN PERFORMED IN ACCORDANCE TO
THE APPROPRIATE IEC 601-1 AND/OR IEC 601-1-1
HARMONIZED NATIONAL STANDARD.
WHEN CONNECTING EXTERNAL EQUIPMENT (e.g., DATA CAPTURE
COMPUTER), THE SYSTEM LEAKAGE CURRENT MUST BE
CHECKED AND MUST BE LESS THAN THE IEC601-1-1 LIMIT.
Note:
When software is upgraded, all monitor configuration settings will be lost.
Additionally, previously recorded data may be lost. Therefore,
configuration settings and previous data should be recorded before the
software update is performed.