SECTION 2 INSTALLATION AND PREPARATION FOR USE
______________________________________________________________________
2-9
The fan prevents temperature inside the monitor chassis from rising above optimum
operating levels.
Caution:
Do not block fan output. Keep at least 3 full inches of unobstructed
space between rear of instrument and wall and other instruments.
Do not allow other instruments to block airflow. Do not block
ventilation inlet holes on the underside of monitor.
The clamp shoe allows the monitor to slide into the pole clamp so that it can be
attached to a ½” – 1 ½” diameter vertical pole.
The potential equalization terminal may be used to ground the monitor and its
components.
The power cord receptacle is used to plug in a three-prong hospital grade power cord
to provide power to the monitor and to the DSC when attached.
The printer port is provided to attach the monitor to the Aspect A-2000 Printer.
Caution:
The A-2000 system complies with the electromagnetic compatibility
requirements of EN60601-1-2. Operation of this device may affect or be
affected by other equipment in the vicinity due to electromagnetic
interference (EMI). If this occurs:
Increase separation between devices
Re-orient device cabling
Plug devices into separate outlet circuit branches
Consult Aspect Medical Systems Technical Support for assistance.
The RS-232 serial port is used to transfer data to and from the monitor. It is also used
in testing and in software updates.
WARNING!
WHEN CONNECTING EXTERNAL EQUIPMENT (e.g., DATA CAPTURE
COMPUTER), THE SYSTEM LEAKAGE CURRENT MUST BE
CHECKED AND MUST BE LESS THAN THE IEC601-1-1 LIMIT.
THE USE OF ACCESSORY EQUIPMENT NOT COMPLYING WITH THE
EQUIVALENT SAFETY REQUIREMENTS OF THIS EQUIPMENT MAY
LEAD TO A REDUCED LEVEL OF SAFETY OF THE RESULTING
SYSTEM. CONSIDERATION RELATING TO THE CHOICE SHALL
INCLUDE:
USE OF THE ACCESSORY IN THE PATIENT VICINITY
EVIDENCE THAT THE SAFETY CERTIFICATION OF THE
ACCESSORY HAS BEEN PERFORMED IN ACCORDANCE TO
THE APPROPRIATE IEC 601-1 AND/OR IEC 601-1-1
HARMONIZED NATIONAL STANDARD.