SECTION 3 OPERATING THE A-2000
______________________________________________________________________
3-13
Time/Date
To set the current time and date:
1. Press [MENU/EXIT] to access the Setup Menu.
2. Highlight Advanced Setup using the up [↑] or down [↓] arrows.
3. Press [SELECT].
4. Highlight Time/Date using the up [↑] or down [↓] arrows.
5. Press [SELECT] until the date or time value you wish to update appears in
reverse video.
6. Use the up [↑] or down [↓] arrows to increase or decrease the values displayed.
7. Press [SELECT] to move to the next date or time value.
8. To save the date or time changes, press [SELECT] at the last Time/Date option
(minutes). Pressing [Menu/Exit] before this last option aborts previous changes.
9. Press [MENU/EXIT] to exit from the Advanced Setup Menu.
Note:
The time/date is initially set for the Eastern Standard Time zone (USA). It
will be necessary for you to change the time twice per year using the
Time/Date feature if you are located in a time zone that alters its clocks at
the beginning or end of Daylight Savings Time.
When a time/date is entered that is earlier than the current time/date in
the monitor, all data memory will be erased. The prompt, "Clear Data
Memory?" will appear to alert you of this. Press [SELECT] to confirm, or
press [MENU/EXIT] to abort.
Print Event - Prints all available data on the most recent marked event.
To print an event:
1. Press [MENU/EXIT] to access the Setup Menu.
2. Highlight Advanced Setup using the up [↑] or down [↓] arrows.
3. Press [SELECT].
4. Highlight Print Event using the up [↑] or down [↓] arrows.
5. Press [SELECT].
Display Parameter Setup – Allows the user to turn filters OFF or ON, change the EEG
sweep speed and amplitude settings, and change the default BIS Log recording interval.
To access the Display Parameter Setup screen:
1. Press [MENU/EXIT] to access the Setup Menu.
2. Highlight Advanced Setup using the up [↑] or down [↓] arrows.
3. Press [SELECT].
4. Highlight Display Parameter Setup using the up [↑
] or down [↓] arrows.
5. Press [SELECT].
The Display Parameter Setup options are discussed in section 3.4.3.
Diagnostic Menu
To display the Diagnostic Menu:
1. Press [MENU/EXIT] to access the Setup Menu.