EasyManuals Logo

Associated Research 620L User Manual

Associated Research 620L
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
26
20. INPUT POWER RECEPTACLE: Standard IEC 320 connector for connection
to a standard NEMA style line power (mains) cord.
21. INPUT POWER SWITCH: Sets the line voltage configuration of the instrument.
In the left position it is set for 115 volt operation, in the right position it is set for
230 volt operation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research 620L and is the answer not in the manual?

Associated Research 620L Specifications

General IconGeneral
BrandAssociated Research
Model620L
CategoryTest Equipment
LanguageEnglish

Related product manuals