PF-HI: The minimum allowable Power Factor that triggers a failure when not
exceeded.
Dwell Time: The length of time that is allowed for the programmed test voltage to be
applied.
Delay Time: The length of time that the programmed test voltage is applied but no
judgment of the set parameters is made. Judgment of the parameters is not made
until the end of the delay time.
Neutral: The status of the neutral relay used during Line Leakage testing. Can be
OPEN or CLOSED.
Reverse: The status of the reverse relay used during Line Leakage testing. Can be
ON or OFF.
Ground: The status of the ground relay used during Line Leakage testing. Can be
set to OPEN or CLOSED.
Meas. Device: The measuring device (MD) selected for the Line Leakage test.
Probe: The points at which the measuring device are placed during the Line Leakage
test. Can be set to: “G-L”, “Ph-L”, and “Ph-Pl”.
Leakage: Leakage measurement setting used during the Line Leakage test. Can be
set to RMS or Peak.
Extended Meters: The metered values displayed during a Line Leakage test. Can be
set to ON or OFF. When set to ON, I-Maximum RMS and MD Voltage RMS are
displayed during a Line Leakage test. When set to OFF, Voltage and Leakage RMS
are displayed during a Line Leakage test.
AC/DC: The metered value of leakage current during a Line Leakage test. Can be set
to AC+DC, AC or DC. When set to AC+DC, the unit measures the composite leakage
current through the MD. When set to AC, the unit measures only the AC component of
the leakage current through the MD. When set to DC, the unit measures only the DC
component of the leakage current through the MD*.
*NOTE: If the AC/DC parameter is set to DC or AC, the minimum delay time is
automatically set to a minimum of 1.3sec.