78
COMMAND
DESCRIPTION
VALUE
FL <file number>
File Load
file number = 1-50
FD
File Delete
Active selected File
FD <file number>
File Delete by
Number
file number = 1-50
FS
File Save
Active selected File
FSA {<file
number>,<file
name>}
File Save-as
file number = 1-50, file name =
Valid ASCII (1)
maximum 11 characters
FM {1|0}
File Memory Link
1=On, 0=Off
FM?
FN {<file
number>,<file
name>}
File New
file number = 1-50, file name =
Valid ASCII (1)
maximum 11 characters
SS <step number>
Step Select
step number = 1-30
ADD
<test,p1,p2,p3…>
Adds all parameters
SAR
Step Add RUN test
SAL
Step Add LLT test
SD
Step Delete
Active selected Step
SD <step number>
Step Delete by
Number
step number = 1-30
SP <prompt
message>
Step Prompt Create
prompt message = Valid ASCII
(1)
maximum 32 characters
SP
Step Prompt Delete
SF {1|0}
Step Fail Stop
1=On, 0=Off
(1) "Valid ASCII" is the character set that is available from the front panel LCD user
interface. Consisting of upper case alphabet (A-Z), numbers (0-9) and decimal point
(.), asterisk (*), dash (-), under bar (_), tilde (~) and space (SP).
FL <memory number>
Load a file by memory number from non-volatile memory into random access memory
RAM.
FD
Deletes the file loaded into RAM from the non-volatile memory space as well as RAM.
A new file must be loaded before and tests can be run.
The “FD” command writes to the non-volatile memory of the
instrument. The non-volatile memory has a limited write cycle
life. Therefore, for programmers who wish to send all
parameters before executing each test, the “FD” command should not be used.