4.3.13. J9 or J10 Connector (I/O inputs/outputs) drawing
1 2 3 4 5 6 7 8 9 10111213141516
a) PLC in NPN mode connection
I
1
(Reset)
Customer
Input boards
16/32 programs
I
2
(Start)
I
3
(Pr 1 + 1)
I
4
(Pr 2 + 1)
Good part
Common
Test fail part
Reference fail part
Alarm
End of cycle
24 V DC
24 V DC
ATEQ or customer
I
6
(Pr 8 + 1)
I
5
(Pr 4 + 1)
I
7
(Programmable input)
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
NPN MODE
Ground
+ 24 V DC
J3
INPUTS OUTPUTS
OUTPUTS
+
-
INPUTS
Customer
0 V
(0,3 A max)
Note: The 24V power supply
must be provided by the
internal power supply of the
ATEQ instrument (0,3A
maximum) OR through an
external power supply provided
by the customer.
In the case of customer
external supply, the ATEQ
instrument can be supply by
the 2 and 4 pins on the J3