Manual, F/T Sensor, Data Acquisition (DAQ) Systems
Document #9620-05-DAQ.indd-20
Pinnacle Park • 1031 Goodworth Drive • Apex, NC 27539 • Tel: 919.772.0115 • Fax: 919.772.8259 • www.ati-ia.com • Email: info@ati-ia.com
54
Figure 4.24—IFPSMC Box FT Connector Labels
Attach the New FT Label
Supplied with the Kit
16. Locate the FT serial number label supplied with the kit and x the label to the
front panel, next to the connector of the recently installed IFPS card. The FT serial
number label should match both the IFPS card and transducer serial numbers
associated with the connector. Refer to Figure 4.24.
17. Connect the male connector on the transducer cable to the connector on the front of
the IFPSMC box corresponding to the recently installed IFPS card.
18. Reconnect the power supply and DAQ cables to the back of the IFPSMC box.
Refer to Figure 4.25.
19. After the procedure is complete, resume normal operation.
Figure 4.25—Power Supply and DAQ Cable Connections
Power Supply Connector
9105-C-PS-IFPSMC
IFPSMC power supply
(Included with Multi-sensor
Interface Box)
Connector 1
to DAQ Card
Connector 0
to DAQ Card
(2) DAQ Cables
(9105-C-SH6868EPM-1)