EasyManua.ls Logo

ATI Technologies Axia80-M20 - Page 36

Default Icon
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual, F/T Sensor, Ethernet Axia
Document #9610-05-Ethernet Axia-09
Pinnacle Park • 1031 Goodworth Drive • Apex, NC 27539 • Tel:+1 919.772.0115 • Fax:+1 919.772.8259 • www.ati-ia.com
36
• The total percentage of the calibrated range used by F
z
and T
xy
axes is greater than 105%. Refer
to the following F
z
T
xy
equation.
Refer to Section 15.3—Calibration Ranges for the calibrated ranges that are used in the
preceding equations.
For example:
An Axia80-M20 sensor that uses calibration range 0 (SI-500-20) is subjected to the following loads
and has the following calibration ranges:
Table 4.8—Example of Force/Torque Out of Range
Axis Applied Load
Calibration Range 0
Table 15.3 Value
F
x
87.5 N 500 N
F
y
-151.6 N 500 N
F
z
-500.0 N 900 N
T
x
1.0 Nm 20 Nm
T
y
2.0 Nm 20 Nm
T
z
-17.5 Nm 20 Nm
The F
xy
T
z
equation simplies as follows:
The F
z
T
xy
equation simplies as follows:
Because the F
xy
T
z
equation simplied to TRUE, bit 30 in Table 4.7 is set.

Table of Contents

Related product manuals