F/T DAQ Installation and Operation Manual
Document #9610-05-1017-06
ATI Industrial Automation, 1031 Goodworth Drive, Apex, NC 27539 USA +1-919-772-0115 www.ati-ia.com
31
Deviation from 22
o
C Typical Gain Error
+ 5
o
C
0.1%
+ 15
o
C
0.5%
+ 25
o
C
1%
+ 50
o
C
5%
Table 6.2—Error Introduced Over Temperature for Non-Gamma Transducers
Deviation from 22
o
C Typical Gain Error
+ 5
o
C
0.1%
+ 15
o
C
0.5%
+ 25
o
C
1.5%
+ 50
o
C
7%
Table 6.3—Error Introduced Over Temperature for Gamma Transducers