84 300-002547-00 rev7
Prepare the Patient for Discharge
Remove the Monarch Bronchoscope System
1. Detach the Bronchoscope System handles from the robot arms and remove
the Monarch Bronchoscope System from the patient.
2. Place it aside or immediately pre-clean and place back in the packaging.
Remove Monarch Navigation Patient Sensors
1. Detach Monarch Navigation Patient Sensors from the patient.
2. Place the Monarch Navigation Patient Sensors on the tower.
Remove the Bronchoscope Patient Introducer
1. Disconnect and discard the Bronchoscope Patient Introducer from the
Bronchoscope Swivel Adapter.
2. Remove the Bronchoscope Swivel Adapter and endotracheal tube from the
patient.
Remove and Clean Equipment
Disassemble and Clean the Monarch Bronchoscope
System
If necessary (e.g. product complaint investigation), the Monarch Bronchoscope
System may be returned to Auris per the following instructions. Otherwise, all
single-use devices, including the Bronchoscope and Bronchoscopic Sheath,
should be disposed of according to applicable national and local laws and
guidelines.
NOTE: Follow hospital, local, and state policies and guidelines for handling
used medical devices.
Pre-cleaning the Monarch Bronchoscope System
1. Wipe down external shafts of the Monarch Bronchoscope and Monarch
Bronchoscope Sheath with pre-cleaning solution.
NOTE: Recommended pre-cleaning solution is First Step by Cygnus (500 mL).
Follow the manufacturer's instructions for specific pre-cleaning solution usage.