Prepare the Monarch Bronchoscope System
WARNING: The bronchoscope and sheath must be lubricated prior to insertion.
Failure to lubricate the bronchoscope and sheath may cause buckling and poor
scope driving responsiveness.
1. Open the Monarch Bronchoscope System package and remove the pouched
tray.
Keep the package and all its components for repackaging, in case the
Bronchoscope System needs to be returned at the end of the procedure.
NOTE: If the bronchoscope and/or sheath is dropped after being removed from
its packaging, replace the bronchoscope and sheath to ensure that its
cleanliness has not been compromised.
2. Open the pouch on the device handle side and pull out the tray, ensuring that
the device handles do not fall out of their wells.
3. Hang the tray on a hook on the side of the Tower.
4. Remove the Monarch Bronchoscope from its packaging tube and lubricate
the shaft. Not lubricating may cause navigation accuracy to be impaired.
CAUTION: Placing lubricant on the camera may result in compromised image
quality.
NOTE: Use only silicone-based lubricants such as Silkospray Universal Silicone
Spray. Do not use water-based lubricants. Use of such lubricants will increase
friction between the bronchoscope and sheath while driving.
5. Insert the Monarch Bronchoscope back into the packaging tube and tray until
the procedure begins.