300-002547-00 rev7 17
Subsystem Monarch Tower Monarch Cart
Max weight on shelf 25 kg
Power Requirement 100-240V~ 50/60Hz 1440W* 100-240V~ 50/60Hz 1000W*
*Requires separate dedicated branch circuits.
The Monarch Platform should be transported, stored, and used in the following
environmental conditions:
Monarch Platform Operating
Non-Operating
(Storage or Shipment)
Temperature 10° C to 25° C -25° C to 70° C
Humidity 30% to 60% Non-condensing 10% to 85% Non-condensing
Noise Generated 63 dB @ 1 meter & 20° C NA
Altitude Up to 3000 meters N/A
Declaration of Emissions
To comply with the regulations on electromagnetic interference for a Class A
FCC Device, all interconnect cables to external devices (for example, monitors)
must be shielded and properly grounded. Use of cables not properly shielded
and grounded may result in the equipment causing radio frequency interference
in violation of the FCC regulations.
All types of electronic equipment may characteristically cause electromagnetic
interference with other equipment, either transmitted through air or connecting
cables. The term EMC (Electromagnetic Compatibility) indicates the capability of
equipment to curb electromagnetic influence from other equipment and at the
same time not affect other equipment with similar electromagnetic radiation from
itself.
Proper installation is required in order to achieve the full EMC performance of the
product. Operate the system with all covers closed. Operating the system with
any cover open may affect EMC performance.
The Monarch Platform may be used safely while connected to other devices if
the devices and their specifications, installation, and interconnection with the
system conform to the requirements of IEC 60601-1 (3
rd
Edition) or IEC 60601-1-
1(2
nd
Edition). This system is suitable for use in the following environment. The
user must assure that it is used only in the electromagnetic environment as
specified.