18 300-002547-00 rev7
Table 1. Guidance and Manufacturer’s Declaration – Electromagnetic Emissions – For All ME
Equipment and ME Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The ME System is intended for use in the electromagnetic environment specified below. The
customer or the user of the ME System should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1 The ME System uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The ME System is suitable for use in all
establishments other than domestic, and may be
used in domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes, provided the following
warning is heeded:
Warning: This equipment/system is intended for
use by healthcare professionals only. This
equipment/ system may cause radio interference
or may disrupt the operation of nearby equipment.
It may be necessary to take mitigation measures,
such as re-orienting or relocating the ME System
or shielding the location.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies