46 300-002547-00 rev7
Camera
1. Plug in the Bronchoscope camera cable to the tower.
2. Verify that the camera is live.
3. Press Navigation to continue.
Navigation
1. Plug in Monarch Navigation Field Generator to the tower while leaving it in
tower cabinet.
2. When the button turns blue, press DONE.
3. Remove Monarch Navigation Patient Sensors from the storage pouch in the
door on the side of the tower.
4. Plug in Navigation Patient Sensors to the tower.
5. Attach Navigation Patient Sensors to the patient using the disposable
adhesive Navigation Patient Patches. Patches are designed for a single
adherence. Attach the labeled sensors as follows:
a. L patient left 8
th
rib mid axillary line.
b. R patient right 8
th
rib mid axillary line.
c. M patient sternal notch.
6. Remove the Monarch Navigation Field Generator Mount from storage
compartment on the side of the tower.
7. Attach the Field Generator Mount to the bed rail on the side of the patient that
the target nodule will be.
8. Remove the Navigation Field Generator from tower drawer.
9. Place the Field Generator into the mount by sliding into the top of it. Make
sure that the notch of the Field Generator slides into one of the holes in the
mount.
10. Position the Field Generator such that it is at the mid-chest wall
approximately 5 cm from the patient’s chest.
11. Swipe left on the Tower screen to move past the navigation setup
instructions.
12. Adjust the Field Generator to ensure that the Patient Navigation Sensors are
within the field of lung volume. On the tower screen, all three sensors will turn
green if the Field Generator is in the optimal position.