auris Monarch Platform - Field Generator or Patient Introducer Mount Clamps Do Not Tighten; Monarch Platform will Not Turn on; Tower and Cart Do Not Turn on Simultaneously

auris Monarch Platform
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
300-002547-00 rev7 99
3. Load and unload the handles from the robotic arms.
4. Reboot the system.
5. If the problem persists, contact Auris Customer Care for support.
Field Generator or Patient Introducer Mount Clamps Do
Not Tighten
1. Detach the mount from the table.
2. Inspect the clamp for fluid residue or debris blocking the clamp. Do not
disassemble the clamp to inspect.
3. Exercise the clamp by loosening and tightening the knobs.
4. If the knobs cannot be tightened, contact Auris Customer Care for a
replacement mount.
Monarch Platform Will Not Turn On
1. Confirm that the Monarch Tower and Cart main power cables are connected
to dedicated power outlets.
2. Confirm that the dedicated power outlets are active.
3. Confirm that the umbilical cable is connected to the Cart.
4. Press the power button on either the Tower or Cart. The light ring around the
power buttons will illuminate when the system is turned on.
5. If the Tower and Cart are connected to active power outlets, the umbilical
cable is connected to the Cart, and the system cannot be powered on using
either the Tower or Cart power buttons, contact Auris Customer Care for
support.
Tower and Cart Do Not Turn On Simultaneously
1. Confirm that the Monarch Tower and Cart main power cables are connected
to dedicated power outlets.
2. Confirm that the dedicated power outlets are active.
3. Confirm that the umbilical cable is connected to the Cart.
If the umbilical cable is not connected to the Cart:
a. Turn off the system.
b. Connect the umbilical cable.
c. Turn on the system.