Administer Saline or Another Agent Through the Working
Channel
If you want to administer cold saline or any other agent through the working
channel with a syringe, you must disconnect the valve on the backend of the
Monarch Bronchoscope and access the working channel with a syringe.
Post-procedure
When the procedure is complete, use the following guidelines when you retract
the Monarch Bronchoscope System.
Retraction
CAUTION: Failure to relax the bronchoscope and sheath prior to retraction may cause
increased stress on the scope.
When retracting the Monarch Bronchoscope, fully relax the scope as you retract.
The Monarch Bronchoscope System will automatically relax the scope if you hold
the Retract joystick down for >25mm or if the Monarch Bronchoscope retracts
into the Monarch Bronchoscope Sheath.
As you retract the Monarch Bronchoscope into the Monarch Bronchoscope
Sheath, the Monarch Bronchoscope will automatically pair with the Monarch
Bronchoscope Sheath and Monarch Bronchoscope System retracts.