auris Monarch Platform - Video (Start Screen) Does Not Appear on Tower and Cart Monitor after System Turns on; Loss of Video on Tower or Cart Touchscreen (During Procedure); Navigation Patient Sensors Not Detected During Setup

auris Monarch Platform
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
100 300-002547-00 rev7
4. Confirm that the light rings around the Tower and Cart power buttons are
illuminated. If only one power button is illuminated, disconnect the umbilical
cable and press and hold the power button on the Tower/Cart to turn off.
5. Once the Tower and Cart are both turned off (power button light rings are not
illuminated), reconnect the umbilical cable to the Cart.
6. Turn on the system.
7. If problem persists, contact Auris Customer Care for support.
Video (Start Screen) Does Not Appear on Tower and Cart
Monitor After System Turns On
1. Confirm that light rings on Tower and Cart power buttons are illuminated.
2. If within a few minutes of turning on the system, the Start screen does not
appear on the Tower and Cart monitors, reboot the system.
3. If problem persists, contact Auris Customer Care for support.
Loss of Video on Tower or Cart Touchscreen (During
Procedure)
1. Confirm that light rings on tower and cart power buttons are illuminated.
2. Remove instruments inserted in the working channel of the Bronchoscope
System.
3. Use the Monarch Controller to relax and then retract the Bronchoscope
System.
4. Disconnect the Bronchoscope System from the robotic arms.
5. Reboot the system.
6. If tower and cart video are not present after system reboot, contact Auris
Customer Care for support.
Navigation Patient Sensors Not Detected During Setup
1. Confirm that sensors are placed on the patient within the Field Generator’s
field of view.
2. Slightly shift the location of the Field Generator to see if the sensor positions
change on the setup screen. If the position of the sensors appears to move
on the setup screen, the patient sensor and field generator positions may
need to be adjusted to bring all sensors into view.
3. Detach and reconnect the Navigation Patient Sensor cable on the Tower
connector.