300-002547-00 rev7 85
2. Immerse distal tip of the Monarch Bronchoscope into the pre-cleaning
solution and aspirate enough pre-cleaning solution to remove gross
contaminants, as you would for standard endoscopes.
3. Aspirate Monarch Bronchoscope in air until dry.
4. Remove the Monarch Bronchoscope Sheath Valve and attach the vacuum
source tubing to the back of the Monarch Bronchoscope Sheath.
5. Repeat steps 2-3 for the Monarch Bronchoscope Sheath.
6. Use gauze to dry the external shafts of the Monarch Bronchoscope and
Monarch Bronchoscope Sheath.
7. Disconnect the Bronchoscope Fluidics Tubing from the Monarch
Bronchoscope and tower and discard the tubing.
8. Unplug and wipe the Monarch Controller before storing it in the tower.
Re-packaging the Monarch Bronchoscope System
1. Unplug the Bronchoscope camera cable from the tower.
2. Insert the device shafts into their respective tubes in the plastic tray and seat
handles into tray wells.
3. Wrap and insert camera cable in the clam shell of the tray.
4. Insert the tray into the red biohazard bag, which is provided on the inner lid of
the shelf box.
5. Fold the end of the biohazard bag 3-4 times and secure fold with the large
labels provided on inner lid of the shelf box.
6. Place the bag snugly in the corrugated nest inside the shelf box.
7. Seal the box on multiple sides with the adhesive strips provided on the inner
lid of the shelf box. Reinforce with packaging tape if necessary.
8. Ship the box via FedEx within 48 hours of procedure. A prepaid return label
has already been placed on the box.
Clean Monarch Navigation Patient Sensors
1. Discard the patient patches used with the Monarch Navigation Patient
Sensors.
2. Unplug the Monarch Navigation Patient Sensors from the tower.
3. Wipe down and stow Monarch Navigation Patient Sensors.