300-002547-00 rev7 57
Item Description
Toggle between Monarch Bronchoscope and Monarch Bronchoscope Sheath.
Irrigation button
Aspiration button
WARNING: Always secure the endoscopic controller. When in the enabled
state, dropping or other unintended joystick motion may lead to accidental
commanded motion which may cause patient injury, such as pneumothorax,
hemorrhage or endoluminal irritation.
Working Channel Instruments
WARNING: Working channel instruments are intended for manual operation
once the bronchoscope has reached the target location. Refer to individual
Instructions for Use included with each working channel instrument for
additional information.
NOTE: Presence of a fluoroscopy system near the Monarch Navigation Field
Generator may compromise navigation accuracy. When using the fluoroscopy
system, you should only refer to the live endoscopic views during manual
biopsy.
NOTE: Metallic working channel instruments near the scope tip may temporarily
compromise navigation accuracy. Examples include: instruments 20mm
proximal to the tip or activation of a radial ultrasound probe within the scope.