300-002547-00 rev7 113
ID Description Message
706 A system error has occurred. All robotic functions
have been disabled. Unload the Bronchoscope
System and reboot the system. If problem
persists, contact Auris support.
707 Camera frame rate issue detected. Check
camera visibility.
708 Camera frame rate issue detected. Check
camera visibility.
709 Failed to initialize camera. A system error has occurred. All robotic functions
have been disabled. Unload the Bronchoscope
System and reboot the system. If problem
persists, contact Auris support.
710 Failed to read image. A system error has occurred. All robotic functions
have been disabled. Unload the Bronchoscope
System and reboot the system. If problem
persists, contact Auris support.
711 Scope disconnected Warning: The bronchoscope cable has been
disconnected. Live Bronchoscope view is
unavailable and may lead to patient injury.
Reconnect the cable and press OK to continue.
712 Image frozen Warning: The bronchoscope image is frozen and
the system has been halted to prevent patient
injury. Press OK to continue and if problem
persists, replace the bronchoscope system or
contact Auris support.
713 Camera FPGA FW or HW
Version Mismatch
A system error has occurred. All robotic functions
have been disabled. Unload the Bronchoscope
System and reboot the system. If problem
persists, contact Auris support.