ID Description Message
900 Fluidics error A system error has occurred and all robotic
functions have been disabled. Press OK to
continue. If problem persists, contact Auris
support.
901 Fluidics Initialization Error A system error has occurred. All robotic functions
have been disabled. To resolve, unload then
reload the Bronchoscope System and click OK to
acknowledge the fault. Then, re-initialize
navigation at the main carina.
902 Fluidics pump error An irrigation pump error has been detected.
Press OK to continue.
903 Fluidics valve error An aspiration pinch valve error has been
detected. Press OK to continue
904 Fluidics max volume Saline delivery threshold has been reached.
Press OK to continue.
1000 Pendant non-recoverable
hardware fault
A system error has occurred. All robotic functions
have been disabled. To resolve, unload then
reload the Bronchoscope System and click OK to
acknowledge the fault. Then, re-initialize
navigation at the main carina.
1001 Controller HW error The endoscopic controller has detected a drop or
upside-down placement and all robotic functions
have been paused. Press the play button on the
controller to make functionality available again.
Press OK to continue.
1002
Pendant Revision Fault
Pendant Revision does not
match
A system error has occurred. All robotic functions
have been disabled. To resolve, unload then
reload the Bronchoscope System and click OK to
acknowledge the fault. Then, re-initialize
navigation at the main carina.