3-16
Chapter 3 Starting and Stopping this Device
(ii) Zero point adjustment during operation
When performing zero point adjustment during operation, see section 3-7.
(iii) Checking measured values
• Using the communicator, check the measured values.
• After checking , disconnect the communication cable, attach the device's case cover, and switch
the process to normal operation.
• If the analog signal output, the displayed values, etc., do not match the process conditions,
recheck the setting range and the like. If this does not solve the problem, carry out the
troubleshooting in Chapter 4, or contact us.
• If the measurement output and display do not stabilize, refer to Chapter 4 and adjust the
damping time constant.
(iv) Cautions following confirmation of readings
After checking the measured values, confirm that the device's case cover is securely closed. If it is
not securely closed, rainwater and the like can enter the device and cause damage to the terminals,
electronics modules, etc., located inside.