98 Maintenance
CFPP Process Analyzer CFPP-4.2 03/2016
CAUTION!
Very cold components and coolant!
Risk of skin damage on direct contact.
FOR THIS REASON:
- Wear protective gloves and protective goggles.
Note: For closing the open coolant lines prepare blind plugs. Suitable blind plugs you
find in the measuring unit enclosure.
For opening the screw connections (2) displace the Amaflex insulating. For absorbing
of coolant prepare a cleaning cloth. Open the screw connections. Close the coolant
lines with the blind plugs.
Open the four locks (4) of the measuring cell housing.
Note: At the bottom side of the cover are located the heat exchanger with peltier
elements and other electrical components. The cover is connected to the
measuring cell housing with a connecting cable via a plug connection at the
bottom of the measuring cell housing.
Lift the cover (3) gently. Separate the connecting cable from the measuring cell
housing by disconnecting the connector at the bottom of the measuring cell housing.
Remove the cover in complete.