94 Maintenance
CFPP Process Analyzer CFPP-4.2 03/2016
Fig. 10.17: Adjusting the outlet pipe
3 Bypass pipe
If the liquid level was below the level mark, screw the outlet pipe (1) further in (one
turn corresponds to 1.25 mm).
If the liquid level was above the level mark, screw the outlet pipe (1) further out (one
turn corresponds to 1.25 mm).
Secure the outlet pipe with the lock nut while holding it in position.
By the same procedure, adjust the bypass pipe (3) so that the distance from the
outlet pipe is about 15 mm.
The liquid level is adjusted.
10.15 Reinstalling the Measuring Cell
!
CAUTION!
The transparent plastic of the measuring cell is not alcohol
-resistant, it will
be damaged by contact with alcohol.
FOR THIS REASON:
- Do not clean the transparent plastic parts with alcohol.