94 Maintenance
CFPP Process Analyzer CFPP-4.2 03/2016
Fig. 10.17: Adjusting the outlet pipe
• If the liquid level was below the level mark, screw the outlet pipe (1) further in (one
turn corresponds to 1.25 mm).
• If the liquid level was above the level mark, screw the outlet pipe (1) further out (one
turn corresponds to 1.25 mm).
• Secure the outlet pipe with the lock nut while holding it in position.
• By the same procedure, adjust the bypass pipe (3) so that the distance from the
outlet pipe is about 15 mm.
The liquid level is adjusted.
10.15 Reinstalling the Measuring Cell
!
CAUTION!
The transparent plastic of the measuring cell is not alcohol
-resistant, it will
be damaged by contact with alcohol.
FOR THIS REASON:
- Do not clean the transparent plastic parts with alcohol.