Maintenance 99
03/2016 CFPP Process Analyzer CFPP-4.2
10.17 Install the cover on the measuring cell housing
Prepare a cover (3 in Fig. 10.20).
The cover is arranged with a connecting cable with a connector.
Plug the connector into the female connector at the bottom of the measuring cell
housing (5).
Position the cover on the measuring cell housing.
Fix the cover on the measuring cell housing using the four locks (4).
CAUTION!
Very cold components and coolant!
Risk of skin damage on di
rect contact. FOR THIS REASON:
- Wear protective gloves and protective goggles.
Close the screw connections (2) of the coolant lines. To this remove the blind plugs.
Reinstall the measuring cell (see page 94).