48 Installation and Disassembly
CFPP Process Analyzer CFPP-4.2 03/2016
7.1.4 Tubing Connections
During installation, product can leak from the lines. Additional detergents and solvents
may be used.
WARNING!
Hazardous substances (product, detergents or sol
vents and their vapors)!
Serious injury or death can result from swallowing, breathing in or skin
contact with harmful substances.
FOR THIS REASON:
-
Before starting work, read the safety instructions, for example in the
safety data sheet for the respective substances.
- Wear your personal protective equipment, at least wear protective
goggles and protective gloves suitable for the respective substances.
- Make sure there is sufficient ventilation.
- Immediately clean up spilled chemicals.
The pipes to which the analyzer is connected contain pressurized liquids and gases.
CAUTION!
Pressurized liquids and gases!
Risk of injury!
FOR THIS REASON:
- Before starting work on components carrying liquid or gas, shut off
the feed lines and ensure that they are depressurized.
- Wear your personal protective equipment, at least wear protective
goggles and protective gloves.
- Do not remove or disable any protective devices.
Before making any pipe connections, install manual shut-off valves on the product feed
line and the instrument air supply line into the analyzer. The shut-off valves must be
installed near the analyzer and must be easily accessible. They must be clearly marked so
that they can be identified as belonging to the analyzer and to the substances they carry.