Initial Operation 59
03/2016 CFPP Process Analyzer CFPP-4.2
8.4 Feeding In Product
Note: The product can be supplied without the analyzer being switched on.
• Make sure that the pressure and temperature in the product feed line correspond to
the specification in the layout drawing and the technical data.
• Make sure that the product feed line is clean (see chapter 8.2).
• Make sure that the sample conditioning system is ready for operation.
• Ensure that the vent-drain system and the recovery system, if installed, are ready for
operation.
• Open the shut-off valve in the product feed line, if it is not already open.
• Make sure that the parameters correspond to the values specified in the supplied
layout drawing and the technical data.
• Product pressure at inlet
• Product flow rate
• Product temperature
The analyzer is now supplied with product.