64 Operation
CFPP Process Analyzer CFPP-4.2 03/2016
9.1 Switching on the Analyzer
To switch on the analyzer, proceed as follows:
Open the shut-off valve in the product feed line, if it is not already open.
Make sure that pressure and flow rates for product and instrument air are adjusted
correctly (see chapter 8).
Check the protective devices (see chapter 2.5).
Switch on the power, using the separator for the power supply (fuse or main switch)
for the analyzer.
The control box and the measuring cell housing are purged with instrument air (loud
noise).
2
1
Fig. 9.1: Overpressure control display
1 Touchscreen
2 Overpressure control display
The remaining purge time is displayed by the overpressure control system (2).