82 Maintenance
CFPP Process Analyzer CFPP-4.2 03/2016
Hazardous substances (product and its
vapors) can leave!
Serious injury can result from swallowing, breathing in or skin contact.
FOR
-
Before starting work, read the safety instructions, for example in the
safety data sheet for the respective substances.
- Wear your personal protective equipment, at least wear protective
goggles and protective gloves suitable for the respective substances.
- Make sure there is sufficient ventilation.
- Immediately clean up spilled chemicals.
• Loosen the blind cape with two size 13 spanners.
• Collect flowing product by using the drain bottle. Wait, until no product is flowing.
The vacuum tank is emptied.
Note: Only in the case of a fault a bigger amount of product accumulate in the
vacuum tank. The fault must be solved as soon as possible. Contact for it
BARTEC BENKE.
• Close the drain professionally with the blind cape (2).
• Dispose the collected product professionally.
• Make sure, that the pressure in the vacuum tank is 20 mbar. The pressure is shown at
the vaccum pressure gauge. It is located in the measuring unit enclosure (see page
Fehler! Textmarke nicht definiert.).
• You can go on with measuring (see chapter 9.2).
10.12 Removing the Measuring Cell
For cleaning the test filter, for checking the liquid level in the sample container and for
changing the cover of the measuring cell housing you must remove the measuring cell.