78 Maintenance
CFPP Process Analyzer CFPP-4.2 03/2016
If this is not the case, increase product flow to the analyzer.
The test filter check is finished.
10.9 Cleaning the touchscreen
To clean the touchscreen, use only water with dish soap or foaming display cleaner as
cleaning agents.
WARNING!
Possible explos
ion or fire!
An explosion or a fire can cause serious injury or death.
FOR THIS REASON:
- Never clean the touchscreen with a dry cloth, but always use a
damp cloth to avoid static electricity discharge and sparks.
- Do not clean the touchscreen with aggressive solvents or abrasives,
not with pressurized air or steam. The touchscreen is a part of the
control box with a protective function, therefore it must not be
damaged.
To clean the touchscreen, proceed as follows:
Switch off the analyzer (see chapter 9.3). This prevents you from operating the
software accidentally while you are touching the touchscreen.
Moisten a cleaning cloth.
Apply some cleaning agent to the cloth, not to the touchscreen.
Clean the touchscreen with the cleaning cloth.
The touchscreen has been cleaned. You can now switch the analyzer back on (see
chapter 9).