86 Maintenance
CFPP Process Analyzer CFPP-4.2 03/2016
!
CAUTION!
The transparent plastic of the measuring cell is not alcohol
-resistant, it will
be damaged by contact with alcohol.
FOR THIS REASON:
- Do not clean the transparent plastic parts with alcohol.
Clean the measuring cell.
The measuring cell has now been removed.
Now you can clean the test filter or you can check the liquid level in the sample container
(see chapter 10.14).
10.13 Cleaning the Test Filter
If you have detected dirt particles while checking the test filter (see chapter 10.8), you can
now clean the test filter as follows:
Remove the measuring cell (see chapter 10.10).
Fig. 10.7: Measuring cell
Wait until the cooling jacket (1) has warmed up to > 0 °C (no ice visible).