Maintenance 85
03/2016 CFPP Process Analyzer CFPP-4.2
CAUTION!
The tips of the temperature sensors, particularly the sample temperature
sensor (glass tip!) are very fragile and can bend or break if handled
carelessly.
FOR THIS REASON:
- Handle temperature sensors with great care.
- If possible, take the sensors from the work area immediately after
removing them, and keep them in a safe place.
• Carefully pull out the temperature sensor.
• Carefully clean them with a soft cloth.
• Keep them in a safe place.
• One after the other, pull off the quick-release couplings of the green (8) and red (7)
pipes from the measuring cell and immediately push them onto the two remaining
plug nipples (3, 2) on the underside of the measuring unit enclosure.
• Pull off the quick-release coupling of the drain pipe (12) and push the drain pipe
slightly to the side.
• Remove any leaked product.
• Disconnect the holder of the decondensation ring from the frame by loosing the two
screws (6). Remove the decondensation ring (11).
• Unscrew and remove the fixing screws (9).
mponents and product!
Risk of skin damage on direct contact.
FOR THIS REASON:
- Wear protective gloves and protective goggles.
- Hold the measuring cell only by its upper transparent plastic parts.
• Place a clean cloth on the surface where the measuring cell will be put down.
• Lift out the measuring cell in an upright position, without tilting it.
• Put down the measuring cell in an upright position.