Maintenance 93
03/2016 CFPP Process Analyzer CFPP-4.2
• Reinstall the measuring cell (see chapter 10.15).
The measuring cell is now ready for operation. You can now switch the analyzer back on
(see chapter 9.1).
10.14.2 Adjusting the Liquid Level in the Sample Container
To adjust the liquid level, proceed as follows:
Fig. 10.16: Lock nut of the outlet pipe
1 Outlet pipe
• Loosen the lock nut (2) of the outlet pipe (1) with needle-nosed pliers while holding
the outlet pipe to keep it in position.