104 Troubleshooting
CFPP Process Analyzer CFPP-4.2 03/2016
Error Possible cause Troubleshooting
Bath temperature is not
reached.
Peltier elements are out of
order.
Change the cover of the
measuring cell housing (see
chapter 10.16). A new
completely converted cover
with heat exchanger and
peltier elements you obtain
from BARTEC BENKE.
Fuse or residual current
circuit-breaker has blown.
Find and remove the cause
of the malfunction.
Thermal fuse of peltier
elements has blown.
Find and remove the cause
of the malfunction.
Temperature of coolant is not
deep enough.
Please refer to the manual
Chiller for Liquids.