28 Technical Data
CFPP Process Analyzer CFPP-4.2 03/2016
Analysis
Measuring range -35 °C ... +10 °C; different specification on request
Comparability
≤ DIN EN / ASTM
Reproducibility
≤ DIN EN / ASTM
Method according to DIN EN 116; ASTM D 6371; IP 309;
BS 2000-309
Measuring cycle discontinuous, cycle time 25 min ... 90 min
Electrical Data
Rated voltage
230 VAC ± 10 % 1Ph.; 50 Hz; different spec. on request
Rated current See type plate, technical documents
Back-up fuse See type plate, technical documents
Protection class IP54 *
Explosion Protection
Marking II 2 G IIC T4 Gb
EC Type-Examination
Certificate
TÜV 09 ATEX 554793
Environmental conditions
Ambient temperature operation 5 °C ... 35 °C
storage -20 °C ... 60 °C
Humidity 5 – 80 % relative humidity, non-corrosive
Sample
Consumption approx. 10 l/h; during purging (approx. 4 min) 40 l/h
Primary pressure 1 bar ... 4 bar
Temperature at inlet 15 °C … 40 °C
Viscosity max. 50 cSt at inlet temperature
Sample quality
filtered 10 µm; water content max. 550 ppm
Auxiliary media
Instrument air
Consumption
Option Vortex
approx. 8 Nm
3
/h (purging phase)
approx. 2.3 Nm³/h (normal operation)
approx. 1.7 Nm³/h in operation (in addition)
Quality According to ISO 8573-1: class 2 or better
Primary pressure 3 bar ... 7 bar