EasyManua.ls Logo

BARTEC BENKE CFPP-4.2 - Page 90

BARTEC BENKE CFPP-4.2
107 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
88 Maintenance
CFPP Process Analyzer CFPP-4.2 03/2016
• Keep a suitable container ready for remaining product from the measuring cell.
• Remove remaining product from the measuring cell.
Fig. 10.9: Removing the test filter
7 Filter assembly
8 Cylinder
• Remove the test filter while holding the filter assembly with one hand by its knurled
part (7).
• Unscrew the cylinder (8) with the other hand.
Fig. 10.10: Test filter
9 Test filter
• Remove the test filter (9).
• Clean the test filter and the sample container with a suitable detergent or solvent.
Disassembly and cleaning are now finished. You can now reassemble the test filter.
7
8
9

Table of Contents