88 Maintenance
CFPP Process Analyzer CFPP-4.2 03/2016
Keep a suitable container ready for remaining product from the measuring cell.
Remove remaining product from the measuring cell.
Fig. 10.9: Removing the test filter
7 Filter assembly
8 Cylinder
Remove the test filter while holding the filter assembly with one hand by its knurled
part (7).
Unscrew the cylinder (8) with the other hand.
Fig. 10.10: Test filter
9 Test filter
Remove the test filter (9).
Clean the test filter and the sample container with a suitable detergent or solvent.
Disassembly and cleaning are now finished. You can now reassemble the test filter.