90 Maintenance
CFPP Process Analyzer CFPP-4.2 03/2016
• Switch off the analyzer and secure it, so that it cannot be switched on accidentally
(see chapter 9.3).
• Close the shut-off valve of the product feed line.
• Remove the measuring cell (see chapter 10.10).
Fig. 10.12: Measuring cell
• Wait until the measuring cell has warmed to about room temperature.
• Always hold the measuring cell upright so that no sample is spilled.
Hazardous substances (product, detergents or solvents and their vapors)!
Serious injury or death can result from swallowing, breathing in or skin
contact with harmful substances.
FOR THIS REASON:
- Before starting work, read the safety instructions, for exam
safety data sheet for the respective substances.
- Wear your personal protective equipment, at least wear protective
goggles and protective gloves suitable for the respective substances.
- Make sure there is sufficient ventilation.
- Immediately clean up spilled chemicals.