EasyManua.ls Logo

Basler DGC-500 - Entering Calibration Mode; Calibration Procedure

Basler DGC-500
154 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Entering Calibration Mode
Calibration is accomplished by performing the instructions called out on each of the HMI calibration menu
screens. The calibration menu screens reside in a branch of Menu 3: Sensing Devices. Figure 5-16
illustrates the process of entering the calibration mode and the following steps describe the process.
1. Navigate to Menu 3: Sensing Devices and scroll up or down to the screen labeled INPUT
CALIBRATE FUNCTION.
2. Perform the steps illustrated by Figure 5-16 to enable calibration (screen indicates FEATURE IS
ON). Use the Select/Enter pushbutton to move right. Use the Raise/Scroll pushbutton to move up.
Figure 5-16. Calibration Mode Navigation
Calibration Procedure
Once calibration is enabled, the following steps are performed to calibrate the DGC-500.
After each calibration step is performed, a screen indicating the success of the previous step is displayed.
Figure 5-17 shows the screens resulting from a successful calibration step and an unsuccessful
calibration step.
Successful Calibration Unsuccessful Calibration
Figure 5-17. Successful and Unsuccessful Calibration Screens
1. While viewing the FEATURE IS ON screen (illustrated in Figure 5-16),
press the Select/Enter pushbutton to begin the calibration process. The
first calibration screen (displayed at right) appears. Apply 240 Vac to the
voltage sensing inputs for phase A (terminal P23) and Neutral (terminal
P30). Press the Lower/Scroll pushbutton.
2. Press the Raise/Scroll
pushbutton. The screen at right is displayed.
Apply 240 Vac to the voltage sensing inputs for phase B (P26) and
Neutral (P30). Press the Lower/Scroll pushbutton.
3. Press the Raise/Scroll
pushbutton. The screen at right is displayed.
Apply 240 Vac to the voltage sensing inputs for phase C (P29) and
Neutral (P30). Press the Lower/Scroll pushbutton.
4. Press the Raise/Scroll
pushbutton. The screen at right is displayed.
Apply 2 Aac to the current sensing inputs for phase A (P10 and P11).
Press the Lower/Scroll pushbutton.
5. Press the Raise/Scroll
pushbutton. The screen at right is displayed.
Apply 2 Aac to the current sensing inputs for phase B (P12 and P14).
Press the Lower/Scroll pushbutton.
UNIT CALIBRATED
RAISE FOR NEXT
CAL SET/UNIT BAD
RAISE FOR NEXT
240 VAC TO A-N
PRESS LOWER KEY
240 VAC TO B-N
PRESS LOWER KEY
240 VAC TO -N
PRESS LOWER KEY
2 AAC TO A CT
PRESS LOWER KEY
2 AAC TO B CT
PRESS LOWER KEY
5-16 DGC-500 Installation 9355400990 Rev H

Table of Contents