3 Disassembly / Assembly
Infusomat® compact
plus
P 1.0 3 - 21
EN
For internal use, only
Disassembly
✓ Unit has been opened.
1. Remove the following parts from the housing, upper part
(
Fig. 3 - 26 for devices with serial number 1199 and lower,
Fig. 3 - 27 for devices with serial number 1200 and higher):
a) Loudspeaker.
(see "Loudspeaker" ➡ p. 3 - 15)
b) Accessory connector.
(see "Accessory connector" ➡ p. 3 - 18)
c) Processor board.
(see "Processor board" ➡ p. 3 - 17)
d) Display board with display (possible for devices with serial
number 1199 or lower, only).
(see "Display board with display" ➡ p. 3 - 19)
Assembly
1. Mount the removed parts to the housing, upper part in reverse
order of removal following the instructions given in the corre
-
sponding sections.
Fig. 3 - 26 Housing, upper part with keypad
Fig. 3 - 27 Housing, upper part with keypad and display