Vital Signs Monitor User’s Manual
Chapter 3-2
3.2 Getting Started
3.2.1 Inspecting the Monitor
1. Before you start to make measurements, carry out the following checks on the
monitor including all connected modules.
——Check for any mechanical damage;
——Check for any incorrect connection of all the external cables and accessories.
2. Plug the power cord into the AC power source. If you are using battery power,
ensure that the battery has sufficient power for monitoring. When you use a battery
for the first time, you must charge it, following the instructions given in Battery
chapter.
3.2.2 Starting the Monitor
Press the power switch, the technical alarm lamp will light up in blue, and then the
physiology alarm lamp lights up in yellow and red, at last goes out. Meanwhile the
system clanks and enters main screen and the temperature module displays screen
information after a sound of “ba”.
Warning
If the monitor is mechanically damaged, or if it is not working properly,
do not use it for any monitoring procedure on a patient. Contact your
service personnel.
Caution
The monitor does not have mains switch. The monitor is switched
completely only by unplugging the power cable from the AC power
source.