V6 Vital Signs Monitor Users Manual
5
Monitor Mode
Patient quantity
1
Trend data
3 kinds resolution: 1Min, 5Min, 10Min.
1Mincan store 96 hours
5 Mincan store 480 hours
10 Mincan store 960 hours
Alarm events
1000
NIBPmeasurement
record
5000
A.4 Measurement Specifications
A.4.1 SpO
2
Digital SpO
2
module
Measurement technic
Digital SpO
2
technic
Monitoring parameters
SpO
2
and PR
SpO
2
Measurement range
Resolution
Accuracy
(0100) %
1%
±2%70100)% SpO2
(069) % unspecified
PR
Measurement range
Resolution
Accuracy
(25255) bpm
1 bpm
±1% or ± 1 bpm, whichever is the greater
SpO
2
alarm range
(0100) %, high/low limit can be adjusted continuously
PR alarm range
(0300) bpm, high/low limit can be adjusted continuously
Nellcor SpO2 module
Measurement technic
Nellcor SpO
2
technic
Monitoring parameters
SpO
2
and PR
SpO
2
Measurement range
Resolution
(0100)%