V6 Vital Signs Monitor Users Manual
6
Accuracy
1%
± 2% ((70100)% SpO
2
) (adult)
± 3% ((70100)% SpO
2
)neonate
± 2% ((70100)% SpO
2
)low perfusion
unspecified (069)%
PR
Measurement range
Resolution
Accuracy
(20300) bpm
1 bpm
±1% or ± 1 bpm, whichever is the greater
SpO
2
alarm range
(0100)%high/low limit can be adjusted continuously
PR alarm range
(0300) bpmhigh/low limit can be adjusted continuously
Masimo SpO
2
module
Measurement range
0%100%
Resolution
1%
Accuracy
70%100%±2%Adult/Pediatric, non-motion conditions
70%100%±3%Neonate, non-motion conditions
70% to 100%±3%motion conditions
0%69%undefined
Average time
2s-4s, 4s-6s, 8s, 10s, 12s, 14s, 16s
PR
Measurement range
25 bpm240 bpm
Accuracy
±3 bpmnon-motion conditions
±5 bpmmotion conditions
Resolution
1 bpm
SpCO
Measurement range
0% to 100%
Accuracy
0% to 40%±3%non-motion conditions
40%unspecified
SpMet
Measurement range
0% to 100%
Accuracy
0% to 15%±1%non-motion conditions
15%unspecified
PI
Measurement range
0.05% to 20%
SpHb