Vital Signs Monitor User’s Manual
17
CO
2
Alarm messages
Cause
Level
CO
2
sensor off
CO
2
sensor is off patient or off the
monitor.
Low
Check airway adapter
CO
2
airway adapter is disconnected with
CO
2
sensor.
CO
2
measurement over
range
CO
2
measurement is over range and need
verify zero.
Low
Zero Required
Sensor or module is not initialized.
Check CO
2
Sampling Line
Sampling tube is occluded or damaged;
Sampling tube is kinked or pinched
Exhaust tube is blocked.
CO
2
sensor error
CO
2
sensor is error.
Sensor no initialized
Sensor or module is not initialized.
Temp
Alarm messages
Cause
Level
Temp Sensorerror1
The temp sensor is error.
Low
Temp Sensorerror2
The temp sensor is error.
Temp OverrangeUp
The measurement value is higher than
43.
Temp OverrangeDown
The measurement value is lower than
34.
Temp Environment
OverrangeUp
The ambient temp is higher than 40.
Temp Environment
OverrangeDown
The ambient temp is lower than 10.
Temp Disconnect
The temp probe is disconnected.
C.3 Prompt Messages
System
Messages
Cause
Level
Recording
Recorder is in printing operation.
No level