V6 Vital Signs Monitor User’s Manual
8
(0 ~ 300) mmHg , high/low limit can be adjusted
continuously
(0 ~ 300) mmHg , high/low limit can be adjusted
continuously
(0 ~ 300) mmHg , high/low limit can be adjusted
continuously
A.4.3 CO
2
(LoFlo)
Capnogram displayed in less than 15 s, at an ambient temperature of
25℃, full specifications within 2 minutes.
Short term drift: ±0.8mmHg over four hours
Long term drift: Accuracy specification will be maintained over a 120
hour period.
Accuracy ( 760mmHg,
temperature is 25℃)
0 mmHg to 40 mmHg, ±2 mmHg
41 mmHg to 70 mmHg, ±5% of reading
71 mmHg to 100 mmHg, ±8% of reading
101 mmHg to 150 mmHg, ±10% of reading
(when RR >80 rpm, all the range is ±12% of reading)
Gas temperature at 25℃
Total system response
time
Sample flow rate and
accuracy
Rate: 50 ml/min
Accuracy: -7.5ml/min~+15ml/min
0%~20.0%(0mmHg~152mmHg ), high/low limit can be adjusted
continuously.