V6 Vital Signs Monitor Users Manual
8
Alarm range
Sys
(0 300) mmHg high/low limit can be adjusted
continuously
Dia
(0 300) mmHg high/low limit can be adjusted
continuously
Map
(0 300) mmHg high/low limit can be adjusted
continuously
A.4.3 CO
2
(LoFlo)
Microstream CO
2
module
Measurement way
Infrared spectrum
Measurement mode
microstream
Warm up time
Capnogram displayed in less than 15 s, at an ambient temperature of
25, full specifications within 2 minutes.
Measurement range
(0~19.7) %0~150 mmHg
Resolution
0.1% or 1mmHg
Stability
Short term drift: ±0.8mmHg over four hours
Long term drift: Accuracy specification will be maintained over a 120
hour period.
Unit
%, mmHg, kPa
Accuracy ( 760mmHg
temperature is 25)
0 mmHg to 40 mmHg, ±2 mmHg
41 mmHg to 70 mmHg, ±5% of reading
71 mmHg to 100 mmHg, ±8% of reading
101 mmHg to 150 mmHg, ±10% of reading
(when RR 80 rpm, all the range is ±12% of reading)
Gas temperature at 25
Total system response
time
3s
Sample flow rate and
accuracy
Rate: 50 ml/min
Accuracy: -7.5ml/min~+15ml/min
Alarm range
0%20.0%(0mmHg152mmHg , high/low limit can be adjusted
continuously.
Alarm accuracy
±0.1% or ±1mmHg
awRR
Measurement range
0 rpm~150 rpm