V6 Vital Signs Monitor Users Manual
9
Measurement accuracy
0 rpm70 rpm: ±1 rpm;
71 rpm120 rpm: ±2 rpm;
121 rpm150 rpm: ±3 rpm.
Alarm range and accuracy
Range: 0rpm~150rpmhigh/low limit can be adjusted
continuously.
Accuracy: ±1 rpm
A.4.4 Temp
Ambient temperature
10~40
Measurement range
34.0~43.0
Measurement part
Ear cavity
Unit
℃,℉
Resolution
0.1℃/℉
Accuracy
Between 35.0~42.0: ≤±0.2
Beyond 35.0~42.0: ±0.3
clinical repeat accuracy
≤±0.3
Measurement time
0.8s
Measurement interval time
12s
Standby time
40s
Memory storage
10
Temp probe alarm
The measured value is beyond the range: Alarm occurs when the
value is lower than 34.0 or higher than 43.0.
The ambient temp is beyond the range: Alarm occurs when the
ambient temp is lower than 10.0 or higher than 40.0.
Fever alarm: Alarm occurs when the measured value is higher than
37.5 and lower than 43.0.