V6 Vital Signs Monitor Users Manual
1
Appendix A Product Specifications
A.1 Safety Specifications
A.1.1 Classification
Classification of Protection against
electric shock
I
Degree of protection against electric
shock
SpO
2
, NIBP, Temp :CF
CO
2
:BF
Degree of protection against hazards
of explosion
Not suitable
Degree of protection against ingress
of liquid
IPX1
Mode of operation
Continuous
Note:
IClass I, internally and externally powered equipment.
CFType CF applied part
BF
Type BF applied part
Not suitable: Equipment is not suitable for use in the presence of flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
A.1.2 Environmental Specifications
Temperature
Operating (540)
Transportation and Storage (-20+55)
Humidity
Operating 15%~85%(noncondensing
Transportation and Storage 10%~93%(noncondensing
Atmospheric Pressure
Operating (7001060) hPa
Transportation and Storage (5001060) hPa