EasyManuals Logo

Biolight V6 User Manual

Biolight V6
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
Vital Signs Monitor User’s Manual
Chapter 6-1
Chapter 6 SpO
2
6.1 Introduction
The measurement of oxygen saturation of arterial blood (also known as pulse
oxygen saturation, usually shortened as SpO
2
) adopts the principles of light spectra
and volume tracing. The LED emits lights with two specific wavelengths, which are
selectively absorbed by oxygenated hemoglobin and deoxyhemoglobin. The optical
receptor measures the changes in the light intensity after the light passes the capillary
network and estimates the ratio of oxygenated hemoglobin and the total hemoglobin.
Wavelengths of the light emitted by the pulse oximeter probe are nominally
660nm for red LED and 940nm for infrared LED.
6.2 Safety Information
Warning
Use only SpO2 sensors specified in this manual. Follow the SpO2 sensors
instructions for use and adhere to all warnings and cautions.
When a trend toward patient deoxygenation is indicated, blood samples
should be analyzed by a laboratory co-oximeter to completely understand
the patient’s conditions.
Do not use the monitor and the SpO
2
sensor during magnetic resonance
imaging (MRI). Induced current could cause burns.
Prolonged continuous monitoring may increase the risk of unexpected
changes in skin characteristics, such as irritation, reddening, blistering or
burns. Inspect the sensor site every two hours and move the sensor if the
skin quality changes. For neonates, or patients with poor peripheral blood
circulation or sensitive skin, inspect the sensor site more frequently.
Check the SpO
2
sensor and its package for any sign of damage before use.
Do not use the sensor if any damage is detected.
SpO
2
% =
×100%
Oxygenated hemoglobin