Vital Signs Monitor User’s Manual
Chapter 13-2
be careful and do not spill liquid onto the instrument and keep any liquid out
of it. When wiping the side panel of monitor, you must be especially careful to
keep water out of all kinds of cable and outlet on the panel.
4. Do not use abrasive material including wire brush or metal brightener during
cleaning because this material will damage the panel and monitor screen.
5. Do not submerge the monitor in liquid.
6. While cable or plug of attachment accidentally gets wet, please rinse it with
distilled water or deionized water and dry it in the environment of temperature
40℃ to 80℃ for at least one hour.
13.3 Cleaning and Sterilizing of Accessories
13.3.1 SpO2 Sensor
The recommended disinfector include isopropyl alcohol 70%, 10% decolourant
solution can be used for sterilization at lower standard. Don’t use undiluted
decolourant (5%~5.25% sodium hypochlorite) or other non-recommended
disinfector in order to avoid damage to sensor.
Do not sterilize sensor by ray, steam or epoxy ethane.
Do not directly submerge sensor in liquid.
To avoid long-time harm to sensor, it is suggested that sterilization to
theproduct be conducted only when necessary according to the regulation of
13.3.2 NIBP Cuff
1. Please regularly clean the product;
2. Remove cuff from connector and pull out airbag from sheath;
3. Submerge clean and soft medical gauze pad or other soft cleaning tools into
fresh water or neutral soapy water, and wring out surplus water from the
submerged gauze then wipe airbag and pipe;
4. Wash the cuff sheath in the clean neutral soapy water;
5. After the sheath and airbag intensive drying, enclose airbag with cuff sheath