Page #64 background imageLoading...
Page #64 background image
Vital Signs Monitor User’s Manual
Chapter 9-4
9.5 Setting Temp
9.5.1 Setting Unit
1. Select TEMP” to enter the temp setup menu.
2. SetUnittoor. The new unit will become effective at the next
monitoring.
9.5.2 Setting Alarm
SelectmenuAlarm Setup to enter the alarm setup menu. Set the alarm
level and alarm limit. The setting value will become effective at the next monitoring.
9.6 Temp Alarm
1. The main unit temperature alarm:
If the measured value is lower than 34.0, an alarm for that the temperature
is lower than the low limit occurs.
If the measured value is higher than 43.0, the alarm for that the temperature
is higher than the high limit occurs.
If the ambient temperature is lower than 10.0, the alarm for that the ambient
temperature is lower than the low limit occurs.
If the ambient temperature is higher than 40.0, the alarm for that the
ambient temperature is higher than the high alarm occurs.
If the measured temperature value is higher than the high limit or is lower than
the low limit, the alarm for that the temperature is high or low occurs.
2. The temperature module alarm:
If the measured value is lower than 34.0, the LCD will display “Lo” and the
buzzer sounds “B-B-B-B-B”;
If the measured value is higher than 43.0, the LCD will display “Hi”, and
the buzzer sounds “B-B-B-B-B”;
If the ambient temperature is lower than 10.0
, the LCD will display “AL”
and the buzzer sounds “B-B-B-B-B”;
If the ambient temperature is higher than 40.0, the LCD will display “AH”
and the buzzer sounds “B-B-B-B-B”;
The temperature module has a fever alarm function: While the temperature is
higher than 37.5, the buzzer sounds “B-B-B”.